Operation Theater Equipment Dealers in Hazaribagh (1 products available)

  • Dry Heat Sterilizer

    Get Price Quote

    Salient Features     Complete GMP model.   Basic structure made from M.S. Heavy angles.   Sterilizer inner walls made from AISI SS 316 and outer walls made from AISI SS 304.   Doors provided on both sides, one opening is in non-sterile area and second opening in sterile area.   ‘A’ grade mineral wool insulation to optimize heat effectiveness and stop heat loses.   Dynamically balanced fan provided for hot air circulation to achieve uniform temperature.   Complete electric panel board with pulley and damper connection.   Process Operation     The Ampoules / Vials placed in trays are washed in non-sterile room and loaded on the trolley. The trolley is moved to the sterilizer on a transfer trolley. Temperature and Times are set on the Digital Controller. The Impeller mounted on the top circulates the hot air horizontally over the charge. The air is heated by elements mounted on the top. Fresh air enters the system through Hepa Filter Module. The system incorporates two doors, one opening into the sterile room and the other opening in the non-sterile room. Both the doors are pneumatically interlocked. The temperature profile during sterilization cycle is recorded on a circular chart recorder, validation of the system is provided by a six-point digital temperature scanner. After the time period required for sterilizing the batch has elapsed, the heating circuit shuts of the pneumatics damper (exhaust) opens and cooling operation commences.