Horizontal Cylindrical Autoclave
Get Price Quote
Located in Tamil Nadu, India, we are actively functioning as a Manufacturer and Supplier of Horizontal Cylindrical Autoclave all over the nation. We are backed by a team of skilled and experienced personnel who makes use of sophisticated machines and latest techniques for their development. Feel completely free to place the order!
Horizontal Cylindrical Autoclave with ISI (BIS 3829) Part I Marking
Inner Chamber : SS 316
Steam Jacket : SS 304
Working pressure : 1.2kg/cm2 & 2.1kg/cm2
Working Temperature : 121°C / 134°C
Door : Single door with Radial locking system
Mode of control : With a multiport valve / Automatic Optional
Pre-vacuum and post vacuum : To eliminate air packs and ensure effective sterilization.
Water protection : In the event of low water the supply to the heaters is discontinued automatically and visually indicated in the control panel.
Pressure switch : An automatic pressure switch adjustable type is provided to set the required pressure.
Vacuum drying : An effective vacuum drying is caused with powerful ejector and vacuum breaker which draws air from the atmosphere. It is filtered, pre-heated(sterilized) and finally enters the working chamber for effective drying.
Water Gauge glass : To note down the level of water glass tube with gauge is provided.
Gasket : Silicon Expansion type
Instrumentation
Safety Valve
Vacuum Breaker
Non Return Valve
Steam Trap
Water inlet and Water outlet Valves
Supply Voltage : 440 volts – 3 phase Sequence of Automatic operation includes the following :
Automatic water filling for the steam generator
Leak test for required time and results with alarm will displayed and data logged.
Bowie-dik test
Spore-test results from approved lab submitted
Pulsing (vacuuming, purging and pre-heating)_ Optional
Sterilization with temp and balance time displayed and data logged
Exhaustion appropriate to liquid, instruments and fabric cycles.
Drying cycle with vacuum pump. – Optional
Final cooling with admission of filtered air (0.2 Micron)
Sterilization completes with F.O values Displayed and data’s logged
Optional :
High Vacuum with vacuum pump
Fully Automatic Micro processor or PLC based programmable with solenoid valve operation. Suitable for Liquid, Instruments and Fabric cycle
Data Logging System
Double Door System
Pulsing Controller